Display 7-lead ECG waveform on one screen, cascade ECG waveform display
Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
Heart rate variability (HRV) analysis function
NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
Drug concentration calculation and titration table functions.
One-touch printing of trend graph
Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
Anti-high frequency surgical unit, defibrillation-proof (special leads are necessary).
CMS 7000 Patient Monitor
Performance Specifications :
ECG
– Lead mode: 3-lead or 5-lead
– Lead selection: I, II, III, aVR, aVL, aVF, V
– Waveform: 5-lead, 2-channel
– 3-lead, 1-channel
– Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
– Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
HR:
– Measurement and alarm range: 15~350bpm
– Accuracy: ±1% or ±1bpm, whichever is greater
– Alarm accuracy: ± 2 bpm
– Resolution: 1 bpm
– ST-segment monitoring:
– Measurement and alarm range: -2.0mV~+2.0mV
– Accuracy: -0.8mv~+0.8mv ±0.04mv or ±10%, whichever is greater
– Other range: unspecified
– Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
– Pacemaker: yes
RESPIRATION
– Method: R-F(RA-LL) Impedance
– Respiration rate:
– Measurement and alarm range: 0~150rpm
– Resolution: 1 rpm
– Measurement accuracy: ±2 rpm
– Alarm accuracy: ±3 rpm
– Apnea alarm: 10~40s
– Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
NIBP
– Method: Oscillometry
– Mode: Manual/AUTO/Continuous
– Interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
– Continuous mode: 5 minutes
– Alarm range: 10 ~ 270 mmHg
– Resolution: 1 mmHg
– Cuff pressure accuracy: ±3 mmHg
– Measurement accuracy:
– Maximal mean deviation: ±5 mmHg
– Maximal standard deviation: 8mmHg
– Over-pressure protection:
– Adult mode: 297±3 mmHg
– Pediatric mode: 240±3 mmHg
– Neonatal mode: 147±3 mmHg
SpO2
– Measurement and alarm range: 0 ~ 100%
– Resolution: 1%
– Measurement accuracy: 70%~100%: ±2%;
– 0%~69%: unspecified
PR
– Measurement and alarm range: 30 ~ 250 bpm
– Measurement accuracy: ±2 bpm or ±2%, whichever is greater
TEMP
– Channel: dual-channel
– Measurement and alarm range: 0 ~ 50℃
– Resolution: 0.1 ℃
– Accuracy: ±0.1 ℃
EtCO2(optional)
– Method: Sidestream or Mainstream
– Measurement range for CO2: 0~150mmHg
– Resolution for CO2:
0 ~ 69 mmHg: 0.1 mmHg
70 ~ 150 mmHg: 0.25 mmHg
– Accuracy for CO2:
0 ~ 40 mmHg: ±2 mmHg
41 ~ 70 mmHg: ±5%
71 ~ 100 mmHg: ±8%
101 ~ 150 mmHg: ±10%
– AwRR range: 2~150 rpm
– AwRR accuracy: ±1 rpm
– Apnea alarm: yes
IBP(optional)
– Channel: dual-channel
– Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
– Measurement and alarm range: -10~300 mmHg
– Resolution: 1 mmHg
– Accuracy: ±2% or 1mmHg, whichever is greater
– Power supply:AC 100-240V, 50/60Hz
– Safety classification: Class I, type CF defibrillation-proof applied part
CMS 7000 Patient Monitor Box Accessories
– Adult fingertip SpO2 probe (5-pin)
– Adult NIBP cuff
– NIBP extension tube
– ECG lead cable
– ECG electrode
– Temperature probe
– Power cord
– User Manual
CMS 7000 Patient Monitor Physical characteristic of
– Dimension: 319 mm(L) × 161 mm(W) × 269 mm(H)
– Weight: 3.6 Kg